Specifications

1. Quick and Precise Analysis: 32-bit embedded processor secures high performance of the system.
2. UK made Electrochemical/Fuel cell Sensor: Highly specificity to alcohol, unaffected by other possible breath contaminants.
3. Data Handling Ability: Compact gas circuit design and advanced analysis process assure fast, accurate and stable testing performance.
4. Active and Passive Blow Methods: Both methods can be applied to provide convenience to the user.
5. Power Supply: Built-in Li-ion battery realises reliable and long-time testing action.
6. Memory Capacity: Data storage of 4096 test results. Any test results can be downloaded to the computer for data backup.

 

Details
Technical data
Principle of measurements
Electrochemical/fuel cell sensor, alcohol specific
Measurement range (BrAC)
0.000mg/L~1.500mg/L
Resolution
0.0001 mg/L
Accuracy: Max. Measure error, in relation to ethanol standard
Ethanol Vapor Concentration
C /(mg/L)
Deviation
C<0.400
±0.020mg/L
0.400≤C<1.000
±5%
C≥1.000
±20%
Repeatability of Indicating Value
Ethanol Vapor Concentration
C /(mg/L)
Deviation
C<0.400
0.007mg/L
0.400≤C<1.000
1.75%
C≥1.000
6%
Calibration
Every 6 months
Temperature Range-Operation
-10℃~45℃
Temperature Range-Storage
-20℃~70℃
Optimum Storage Condition
Temperature 20℃±5, Humidity: <80%RH, Pressure: 86kPa~106kPa, No vapor or electromagnetic disturbances in the environment are influencing the normal working conditions of the device.
Blowing Time
2.5 seconds (breath flow≥20L/min)
Warm-up Time
Less than 5 seconds after switching on
Recovery Time
5 seconds to 5 minutes, depends on Ethanol Vapor Concentration
Analysis Time
Less than 10 seconds
Display
128×64 dot matrix black and white screen
Dimensions
Approx. 127mm×60mm×28mm
Memory
4096 records
Operation Voltage
State of Charging
State of Operation
5 V
DC 3.6~4.2V
Battery
Built-in 1000mAh/3.7V Li-ion polymerized battery
Weight